|


What
is PhaseMode?
PhaseMode
is to WaveMode as Lateral Force Mode (LFM) is to Contact Mode. That
is to say, LFM is another type of scan mode that can be acquired
while running the AFM in contact mode (see
Applications Note AN-4). The signal for LFM is generated
when running in contact mode, whether or not the data is being
collected and plotted as an image. Similarly, when running in
WaveMode using amplitude feedback with a fixed resonant frequency, a
phase-shift signal is generated and this can produce image contrast
- and we call this scan mode PhaseMode. The exact nature of the
interaction mechanism that creates the phase shift signal is
uncertain; however, one mechanism almost certainly involves sticking
of the probe to the surface, causing a delay time in the cantilever
resonance pattern, relative to before the instant that the tip makes
its brief contact.
Why
bother with PhaseMode?
There
has been a lot of current interest in PhaseMode. Much of the work
has been done with heterogeneous mixtures of polymers, which tend to
respond well to PhaseMode because the modulus properties of the
polymer are readily resolved. PhaseMode can yield results that are
similar in appearance to LFM, however, PhaseMode is generally more
sensitive in producing meaningful image contrast. Furthermore,
PhaseMode seems to produce image contrast on a greater variety of
sample types than does LFM.
Topography & Material Properties that Control Image Contrast
Subtle
topographic features are well resolved with PhaseMode, even better
than WaveMode, because these features can produce a strong
phase-shift signal. The phase image contrast that results from
topographic features has the appearance of a 2-D derivative image,
similar to that produced in mapping the tip deflection signal (i.e.,
error signal), and can be termed a pure phase image, because it is
not dominated by the effects of material properties. Figure 1 shows
a pure phase image dominated by topographic features.
In addition to topographic features, there are two material
properties that should affect the magnitude of the phase shift
signal when operating PhaseMode at a high resonance amplitude and a
high degree of damping (i.e., hard tapping). First, the degree of
sticking between the tip and sample will be related to the chemical
affinity for bonding between these two objects. Second, the amount
of sample deformation at the surface, as a function of elastic
modulus, will affect the contact area between the tip and the
sample. These two effects are synergistic, because the more contact
area there is between the tip and the sample, the greater the degree
of sticking. Figure 2 is a PhaseMode image that was produced using
hard-tapping, and consequently shows a map of force-modulus .

Control
Variables:
After
careful systematic study, it is clear that a variety of experimental
parameters can be used to control image quality in PhaseMode. It is
fair to say that there is not necessarily a bad or good image
produced with PhaseMode, but rather, some images have contrast
characteristics that seem more useful than others. The challenge,
therefore, is to control experimental parameters so as to create
images that give the best contrast. The following are experimental
parameters that used for getting "good" phase mode images:
1.
Amplitude damping (%) - for pure phase, damping is set <
50%; for force modulus, damping is set < 65%
2. AC drive voltage (V) - for pure phase, max peak
intensity < 1/2 scale; for force modulus, max peak intensity
> 1/2 scale
3. Tip radius - a partially dulled probe tip is better for
more contrast in force modulus imaging
1.
Amplitude damping (%) - If the % damping is too high, that is,
if the amount of amplitude damping is very high (i.e., a small
negative set-point), then the image contrast can be either too
great, or sometimes there is no contrast at all (Figures 3a-3d).
Conversely, if the % damping value is too small (i.e., a large
negative set-point), then the tip will not properly track the
surface because there is not enough energy transferred from the
probe to the sample for a good feedback response. There is a range
of damping values that produce reasonably good results, varying from
30% to 85%. After setting the %-damping value as prescribed, one can
"fiddle" around with this parameter in order to optimize
contrast features
2.
AC drive voltage - The effect of voltage is controlled in a way
similar to resonance damping %. If the resonance voltage is too
high, poor contrast results. If the voltage is too low, then
artifacts creep into the image as a result of the tip not properly
tracking the surface (this is true for WaveMode in general). The
effects of varying voltage can be seen in a series of images taken
from a piece of rubber tire (Figures 4a-1f). At a lower voltage, the
image is that of a purely derivative nature (Figure 4a). With
gradually increasing voltages, the images become more dominated by
the effects of sample modulus (i.e., Figure 4f shows no remaining
pure phase signal).
3.
Tip radius - Recall, the amount of contact area between the tip
and sample will control the amount of sticking between these
objects, which in turn controls the amount of phase lag. Consider
that not only will the elastic modulus of a material control the
contact area, but the radius of the probe tip is also an important
factor. A sharp probe tip is often less than ideal for getting good
force modulus contrast with PhaseMode - instead, a dull probe tip
often seems to yield a better response. However, a better pure
PhaseMode image is obtained with a sharper probe tip simply because
topographic features are better resolved with the smaller
interaction area of a smaller tip radius. One downfall to this
reality is that there exists a point where a probe tip is damaged
beyond any usefulness. That is when the probe tip has been dulled so
much that image artifacts are produced. At this point, the probe
should be discarded.
|