|
Scanning probe microscopes universally use PZT material either in
the form of cylinders or as stacked elements to position the probe
in the X, Y, and Z axes. PZT is a lead zirconate titanate
piezoelectric ceramic. The most serious limitation of this material
is its tendency to continue to expand after a voltage is applied to
it. This phenomenon is generally called Creep or Drift, and
continues to occur over at least seven orders of magnitude in time.
While the degree of drift may be only a few percent in one decade of
time, it makes the difference between a precision instrument and one
that is useful only for crude measurements. The manufacturers of the
PZT do not provide much information about creep, but may show it as
a graph where the PZT extension increases linearly with the
logarithm of time. Our measurements have shown this to be only a
crude approximation. The amount of creep and time dependence will
vary with each type or formulation of PZT material.
The artifacts caused by creep are present in each axis. The Z axis
responds to signals from seconds to milliseconds. The fast scan axis
responds to signals from seconds to tenths of a second, and the slow
axis responds to signals from hundreds of seconds to seconds. In the
X and Y axis, the calibration between the axes will generally differ
by 20% with the slow axis having a greater movement. Rotation of the
direction of scan by 90 degrees will therefore reverse the
difference and cause a 40% distortion in the scan. The first scan
will not agree with subsequent scans since the PZT has a memory of
at least 5 minutes. The creep causes a phase shift between the
Fourier components of the sweep waveforms, resulting in a highly
nonlinear scan. The nonlinear correction and calibration change with
all parameter of the scan, such as sweep speed, image resolution,
and direction. In the Z axis the effect is most obvious in the shift
of the baseline and the inability to obtain accurate calibration.
Software compensation is not practical due to the long memory of the
PZT.
An obvious way to deal with the problem is to use an independent
means of sensing the position of the PZT and to use whatever voltage
is needed to push the PZT to the right position. There are a number
of problems with this idea. If strain gauges are used, the full
scale output is typically about 5 millivolts. Obtaining a noise free
signal is possible only at fairly large scans, and at low speeds.
Capacitive sensors provide a larger output, but add considerably to
the system cost, and make the system much more delicate, trouble
prone, and may require a factory technician to calibrate the system
in the field. Usually, for small scans, the closed loop system is
switched off. In any event, the more precise the system is before
correction, the better it will be after.
PZT material is not homogeneous, but consists of millions of grains,
and within these grains are domains with random orientation. The
capacitance of these grains will vary throughout the material. The
resistivity of the material is also not infinite. Like semiconductor
materials, the Fermi level lies below the conduction band, and at
any temperature above absolute zero, there are thermally excited
electrons that are free to move. As a result of this, current will
flow through the material in response to local potential gradients.
When a voltage is applied to the PZT, the potential gradients are
immediately determined by the capacitive reactance of the material,
and an initial movement occurs. In the long term, the potential
distribution will be determined by the bulk resistivity of the
material, which is not directly related to capacitance. Creep is
caused by the relaxation process whereby the equipotential surfaces
adjust themselves to conform to the resistivity of the material.
There are short paths of current flow that occur in milliseconds,
and much longer paths that require many minutes.
In an electronic analogy, the PZT is like a large mesh of resistors
and capacitors connected between two terminals. The values of the
elements vary in a random fashion. The relaxation is a linear
phenomenon and therefore proportional to the applied voltage. The
I-V response is completely specified by the impulse response, or its
integral the step response. The response to an arbitrary voltage
waveform is given by the convolution of the PZT response and the
applied voltage.
In addition to providing a metrology head with precise position
sensing, a solution pursued at Quesant with every system is to
provide electronic compensation networks in series with the PZT
drivers that equalize the response over the whole range of time
constants, from 5 millisec. to 300 seconds. This effectively makes
the PZT material respond to any input signal in an ideal way. A step
function for instance results in a step movement with no creep.
Software developed for the purpose allows adjustment of each system.
With compensation, the hard zoom is free from drift, step
calibration gratings have flat tops on the steps, and the
calibration does not depend on the orientation of the grating.
Nonlinearity of the scans is dramatically reduced and does not
depend on the direction of the scan.
Before purchasing a probe
microscope, tests should be performed to see how it will work on
meaningful samples:
1.
Measure a calibration grid, then rotate the scan direction 90
degrees to see if the calibration holds. Keep the grid fixed.
2. Locate a blemish and position it near the corner of the
scan. Do a hard zoom and look for distortion or drift of the
blemish on subsequent scans.
3. Measure a step height grating in one orientation and
rotate it 90 degrees. If the measurement is not the same, or the
baseline is not straight, there is creep in Z.
4. Measure a precise grating like a ruled diffraction
grating in 0 degrees and 90 degree orientation. There should be no
noticeable change in the line spacing to the eye.
5. Make a slow scan of a diffraction grating (0.5 Hz) and
then a fast scan (20 Hz). The images should be exactly the same in
calibration and linearity.

|